-30°C Freezer Calibration|7-Point Temperature Mapping & Panel Comparison

This calibration was conducted on a -30°C laboratory freezer, commonly used for sample preservation and reagent storage. To ensure reliable performance in low-temperature conditions, READY performed the following tests:

7-Point Temperature Mapping

  • Installed 7 calibrated probes across different freezer zones (top, middle, bottom, and corners).

  • Recorded data every 5 minutes for 12 hours (144 readings per point).

  • Analyzed temperature uniformity and stability across all test points.

Panel vs. Standard Thermometer Comparison

  • Used a traceable, calibrated thermometer for reference measurements.

  • Compared actual readings against the freezer’s panel display to verify accuracy.

Calibration Report

  • Delivered complete data curves, statistical analysis, and deviation evaluation.

  • Report can be used for internal quality assurance or external audits.

READY ensures your freezer operates with high precision and stability, safeguarding sample storage conditions.

Looking to verify if your -30°C freezer meets laboratory quality standards? Need an audit-ready calibration report?