For biological sample storage, cell banking, clinical specimens, and long-term reagent preservation, the stability of an -80°C Ultra-Low Temperature Freezer is crucial. READY Biotech provides complete calibration services, including temperature mapping, display comparison, long-term monitoring, and uniformity analysis to meet QA and audit requirements.


1-Display Temperature vs Standard Reference Thermometer

A traceable reference thermometer is used to verify freezer temperature accuracy.

Includes:

  • Display temperature vs. true standard reading

  • Verification of setpoint accuracy

  • External sensor comparison (if available)

This detects deviations, probe drift, or unstable temperature control.


2-24-Hour Temperature Logging (Every 5 Minutes)

To evaluate long-term stability:

  • Logging interval: every 5 minutes

  • Test duration: 24 hours

  • Deliverables: raw temperature data + trend graph

This confirms:

  • Stability over long operating periods

  • Compressor cycling behavior

  • Day/night temperature variation


3-Three-Point Temperature Mapping (Top / Middle / Bottom + Sensor)

Sensors are placed at multiple levels to assess uniformity:

  • Top

  • Middle

  • Bottom

  • Built-in sensor (if applicable)

Used to identify hot/cold zones and validate freezer performance under QA standards.


4-Report Includes

  • 3-point temperature graphs

  • 24-hour raw data

  • Display vs standard thermometer comparison

  • Uniformity analysis

  • Recommendations when deviations occur