HIRAYAMA HV-85 Autoclave Validation|Heat Distribution, Heat Penetration & BI Test


Autoclave validation is essential to ensure effective and consistent sterilization performance.
This validation service for HIRAYAMA HV-85 autoclave includes heat distribution, heat penetration, biological indicator testing, and pressure testing to confirm that the autoclave achieves sufficient sterilization temperature and time under defined cycle conditions.


? Validation Scope

  • Heat Distribution Test (Empty Load)
    121°C for 15 minutes, 4 temperature measurement points

  • Heat Penetration Test (Full Load)
    121°C for 30 minutes, 4 temperature measurement points

  • Biological Indicator (BI) Test
    BI placement and incubation to verify sterilization efficacy

  • Pressure Test
    Monitoring pressure behavior during the sterilization cycle


? Validation Report

The final report includes:

  • Temperature data records

  • Temperature and pressure profiles

  • F0 value calculation

  • BI incubation results