Autoclave Validation|Heat Distribution & Penetration Test
Autoclave Validation|Heat Distribution & Penetration Test|READY Biotech
Autoclaves are essential for laboratory sterilization. To ensure effective sterilization performance, READY Biotech provides comprehensive autoclave validation services, including:
▣ Test Items
-
Heat Distribution Test (Empty Chamber) – 3-point temperature measurement at 121°C to verify uniform temperature across the chamber.
-
Heat Penetration Test (Loaded Chamber) – Validate sterilization effectiveness under full load, ensuring steam penetration.
-
F0 Value Verification – Confirm accumulated F0 value meets the required kill value during sterilization.
-
Biological Indicator (BI) Test – Place BI strips (e.g., BIONOVA) inside the chamber and incubate to confirm sterilization success.
▣ Service Assurance
-
All reference standards are calibrated by TAF-accredited third-party laboratories.
-
Reports include detailed temperature curves, suitable for internal audits or GMP/GLP inspections.
-
Custom validation protocols can be tailored according to client SOPs.