Autoclave Validation Service|Heat Distribution, BI Test & Timer Verification

Autoclave Validation: Ensuring Reliable 121°C Sterilization

READY Biotech provides full autoclave validation including empty chamber heat distribution, load heat penetration, biological indicator (BI) test, pressure test, and 31-minute timer verification, with complete F0 values and temperature curve reports.


01|Empty Chamber Heat Distribution

  • Temperature sensors placed throughout the empty chamber

  • Verifies uniform heating to 121°C

  • Identifies cold spots that may reduce sterilization effectiveness


02|Load Heat Penetration Test

  • Performed with simulated or customer-specified load

  • Confirms temperature penetration to the load’s core

  • Ensures real-world sterilization reliability


03|Biological Indicator (BI) Test

  • Uses Geobacillus stearothermophilus spore indicators

  • Post-cycle incubation and observation

  • Pass/Fail determination based on microbial growth

  • Full documentation with photos


04|Pressure Test (Equipment Dependent)

Performed only when the autoclave model allows:

  • Checks pressure rise and release behavior

  • Compares actual vs. displayed pressure

  • Identifies leaks or seal issues


05|Timer Verification – 31 Minutes at 121°C

Once the autoclave display stabilizes at 121°C:

  • Start both the built-in timer and a certified reference timer

  • Compare the full 31-minute sterilization holding time

  • Ensures accurate exposure time for the sterilization cycle


06|Comprehensive Validation Report

Includes:

  • Temperature curve graphs

  • F0 value

  • Heat distribution / penetration data

  • BI test results

  • Pressure test (if applicable)

  • Timer verification result