Autoclave Validation Service|Heat Distribution, BI Test & Timer Verification
Autoclave Validation: Ensuring Reliable 121°C Sterilization
READY Biotech provides full autoclave validation including empty chamber heat distribution, load heat penetration, biological indicator (BI) test, pressure test, and 31-minute timer verification, with complete F0 values and temperature curve reports.
01|Empty Chamber Heat Distribution
-
Temperature sensors placed throughout the empty chamber
-
Verifies uniform heating to 121°C
-
Identifies cold spots that may reduce sterilization effectiveness
02|Load Heat Penetration Test
-
Performed with simulated or customer-specified load
-
Confirms temperature penetration to the load’s core
-
Ensures real-world sterilization reliability
03|Biological Indicator (BI) Test
-
Uses Geobacillus stearothermophilus spore indicators
-
Post-cycle incubation and observation
-
Pass/Fail determination based on microbial growth
-
Full documentation with photos
04|Pressure Test (Equipment Dependent)
Performed only when the autoclave model allows:
-
Checks pressure rise and release behavior
-
Compares actual vs. displayed pressure
-
Identifies leaks or seal issues
05|Timer Verification – 31 Minutes at 121°C
Once the autoclave display stabilizes at 121°C:
-
Start both the built-in timer and a certified reference timer
-
Compare the full 31-minute sterilization holding time
-
Ensures accurate exposure time for the sterilization cycle
06|Comprehensive Validation Report
Includes:
-
Temperature curve graphs
-
F0 value
-
Heat distribution / penetration data
-
BI test results
-
Pressure test (if applicable)
-
Timer verification result