Biosafety Cabinet BSC Calibration Record | Complete Inspection of Nine Test Items
Calibration Record|Biological Safety Cabinet (BSC) Performance Test|READY Biotech
To ensure your Biological Safety Cabinet (BSC) continuously provides personnel, product, and environmental protection, READY Biotech conducted a complete 9-item field calibration and performance test based on NSF/EN standards.
☛ Calibration Items:
▪ Inflow Velocity Test
Measures front airflow velocity to verify containment effectiveness.
▪ Downflow Velocity Test
Checks airflow uniformity within the work zone to ensure sample protection.
▪ HEPA Filter Integrity Test (PAO / Aerosol Leak Test)
Ensures no leakage through the filter media or housing.
▪ Smoke Pattern Test
Visual inspection of airflow direction and turbulence using smoke.
▪ UV Intensity Test
Checks if the UV sterilization light is emitting sufficient irradiance.
▪ Illumination Test (Lux)
Verifies that lighting level is adequate for safe operation.
▪ Noise Test
Measures sound level during normal operation to ensure compliance.
▪ Vibration Test
Assesses platform vibration to ensure suitability for sensitive work.
▪ General On-Site Evaluation
Covers panel function, filter condition, visual inspection, and user guidance.

☄ ︎ Annual calibration is recommended, especially for laboratories under GLP, TFDA, or ISO/IEC 17025 requirements.
☄︎ Provide your cabinet model and brand—we'll assess and confirm compatibility for calibration. All brands are welcome, not limited to Thermo or ESCO!