Dual Temperature Incubator Calibration|Upper 32.5°C & Lower 22.5°C 10 Points / 1 Hour
Dual Temperature Incubator Calibration|Upper 32.5°C & Lower 22.5°C|10 Points / 1 Hour|READY Biotech
The calibrated instrument is a dual temperature incubator, commonly used for cell culture, sample preservation, and temperature-sensitive experiments. Since it operates with independent temperature control for upper and lower chambers, calibration is required to ensure accuracy and stability.
READY Biotech performed the following validation procedures:
▣ Multi-Point Temperature Distribution Test
-
Upper chamber set at 32.5°C; lower chamber set at 22.5°C.
-
Installed 10 calibrated temperature sensors in each chamber, covering central and peripheral zones.
-
Temperature recorded every 5 minutes for 1 hour (12 data points per sensor).
-
Analyzed point-to-point variations and fluctuations to confirm uniformity and stability.
▣ Panel Display Verification
-
Compared the incubator panel readings against a traceable calibrated temperature probe.
-
Verified deviations were within acceptable tolerance limits.
▣ Calibration Report
-
Delivered complete data curves and statistical analysis.
-
Report can be used for internal quality assurance, regulatory audits, and GMP/GLP compliance.
READY calibration services help ensure dual temperature incubators perform reliably, supporting consistent and reproducible experimental outcomes.