Dual-Temperature Refrigerator Calibration|4 / −20 °C Multi-Point Temperature & Alarm Verification
Dual-Temperature Refrigerator Calibration|4 / −20 °C Multi-Point Temperature & Alarm Verification|READY Biotech
In biological storage, vaccine facilities, and diagnostic laboratories, dual-temperature refrigerators are critical. They maintain both refrigeration and freezing zones. Any deviation in temperature control or alarm failure can jeopardize stored samples. READY Biotech offers professional calibration and testing services to ensure both 4 °C and −20 °C zones remain accurate and reliable.
? Test Items & Procedure
-
12-Point Temperature Mapping (12-Hour Recording)
• Deploy 12 sensors across both refrigeration and freezing zones (top, middle, bottom; front, center, back)
• Record continuously for 12 hours, logging data every minute or few minutes
• Analyze temperature distribution, fluctuations, and detect any hot/cold spots -
External Electronic Data Logger Verification
• If the client has an external electronic temperature recorder (datalogger), we parallel it with our standard sensors for comparison
• Validate the logger’s accuracy, stability, and calibration deviation -
Alarm System Testing
• Simulate temperature excursion scenarios above or below set limits in both zones to verify audible and visual alarm triggering
• Test alarm response time and external alarm outputs -
Reporting & Deliverables
• Provide a full calibration report with all 12-point temperature trend charts, statistical analysis, and alarm test results
• Report is suitable for internal quality management or audit submission