Primary Filter Replacement for Laminar Flow Cabinet | READY Biotech

In a laminar flow cabinet, air cleanliness directly affects experimental accuracy. READY Biotech provides primary filter replacement services, evaluating model compatibility and ensuring optimal performance after replacement.

? Difference Between Primary and HEPA Filters

A laminar flow cabinet typically includes two filtration stages:

  • Primary Filter (Pre-filter): Installed at the air inlet to remove large particles and fibers, protecting the HEPA filter from clogging. Recommended replacement every 3–6 months depending on usage.

  • HEPA Filter: Installed at the exhaust end, removing ≥0.3μm particles with 99.97% efficiency to maintain ISO Class 5 clean air. Typically replaced annually after professional airflow testing.

A clogged primary filter reduces airflow speed and increases pressure across the HEPA filter, risking contamination. Regular replacement maintains efficiency and prolongs HEPA filter life.

? READY Service Process

  1. Evaluate replaceability based on brand/model (ESCO, Thermo, Yamato, etc.)

  2. Remove old filter and clean the filter frame

  3. Install new primary filter and check sealing

  4. Restart the system and verify airflow balance

  5. Provide a replacement report with next maintenance reminder