Floor-standing High-Speed Refrigerated Centrifuge Verification Service

List of Laboratory Centrifuge Operating Condition Testing ItemsREADY

Floor-standing high-speed refrigerated centrifuges are widely used in biotechnology, pharmaceutical, and research laboratories.
The accuracy of rotational speed, operating time, and chamber temperature directly affects sample separation performance and experimental reproducibility.

Regular verification ensures that the centrifuge operates according to the set parameters and provides reliable documentation for internal quality management and audit purposes.

The verification scope includes:

Speed Verification
Verification range from 300 to 14,800 rpm. Actual rotational speed is measured under different setpoints and compared with calibrated reference instruments.

Time Verification
Verification range from 1 to 30 minutes. Actual operating time is measured using a calibrated timer to confirm accuracy against the displayed value.

Temperature Verification
Temperature verification is performed at setpoints of 4°C and 20°C. After stabilization, actual chamber temperature is measured and compared with calibrated temperature reference devices.

Verification results can be used for equipment performance confirmation, quality records, or audit documentation. Acceptance criteria should be defined according to the user’s SOPs and operational requirements.