This record covers the calibration of a low temperature incubator, commonly used for cell culture, microbiology, and pharmaceutical storage. To ensure accuracy and stability, READY Biotech performed the following calibration / testing / validation procedures:

Multi-Point Temperature Distribution Test

  • Conditions set at 20°C, 40°C, and 60°C

  • Six temperature sensors placed across upper, middle, and lower levels

  • Data recorded every minute for 30 minutes after stabilization

Panel Display Comparison & Accuracy Verification

  • Measurements taken with a calibrated reference thermometer

  • Compared incubator panel display with actual measured values

  • Confirmed deviations within acceptance criteria

Reporting & Compliance

  • Full data curve and stability analysis provided

  • Report structured to meet GMP / GLP compliance requirements

READY’s calibration and testing service ensures your low temperature incubator performs reliably, safeguarding experiment reproducibility and regulatory compliance.