Oscillating incubator calibration record | Temperature and speed test
Calibration Record|Shaking Incubator Speed & Temperature Validation|READY Biotech
To ensure the reliable performance of the shaking incubator during microbial and cell culture applications, this calibration session focused on validating the system's temperature and shaking speed accuracy under typical operating conditions.
✎Calibration Items & Results:
▪ Temperature Validation
Three temperature sensors were placed inside the chamber (top, middle, bottom) to record a 30-minute continuous temperature profile. All readings were within ±1.0°C of the setpoint.
▪ Shaking Speed Verification
A digital tachometer was used to measure the actual shaking speed. Results showed that rpm deviation was within ±5 rpm, confirming stable and accurate platform movement.
✎Notes & Recommendations:
▪ Calibration results meet the defined acceptance criteria.
▪ Annual calibration is recommended to maintain reliable performance, especially for high-precision or high-frequency applications.