4°C Refrigerator Temperature Validation|15-Point Mapping, 24-Hour Study
4°C refrigerators are critical for storing biological samples, reagents, and pharmaceuticals. Uneven temperature distribution may compromise sample integrity and regulatory compliance.
READY Biotech provides comprehensive temperature mapping under both empty and loaded conditions based on customer SOP and QA requirements.
Test Conditions|Empty & Loaded Studies
-
Empty Load Test: Evaluates intrinsic temperature control performance
-
Loaded Test: Simulates actual storage conditions
15-Point Temperature Mapping Design
Temperature sensors are positioned to cover:
-
Top, middle, and bottom shelves
-
Front, center, and rear zones
-
Potential hot and cold spots
Data Logging & Monitoring Duration
-
Logging Interval: 1 data point per minute
-
Monitoring Duration: Continuous 24 hours
This allows assessment of long-term stability and operational impact.
Report Deliverables
The validation report includes:
-
Complete raw temperature data (15 points)
-
Temperature trend curves
-
Comparison of empty and loaded conditions
-
Acceptance evaluation based on customer criteria
☎︎ Tel: 02-8228-0131