Shaking incubators are essential equipment for microbial culture, cell growth, enzyme reactions, and molecular biology applications. Stable temperature control and consistent shaking are critical for maintaining experimental accuracy. Deviations may directly affect culture quality and reproducibility.
READY Biotech offers professional shaking incubator calibration, featuring 3-point internal temperature monitoring with continuous recording for 1 hour to ensure uniformity, stability, and reliable temperature performance.
? Test Items & Procedure
3-Point Temperature Monitoring (1-Hour Continuous Recording)
• Deploy three temperature sensors at key positions inside the incubator (top/middle/bottom or front/center/rear)
• Recording duration: minimum 1 hour
• Data logging interval: every minute or several minutes
• Analysis includes:
-
Temperature uniformity (spatial variation)
-
Stability over time
-
Detection of potential hot or cold spots
This test provides a clear evaluation of the incubator’s temperature performance during actual operating conditions.
Operational Status Check
• Compare set temperature vs. actual temperature
• Evaluate heating stabilization and temperature behavior during shaking operation
• CO₂ or humidity control (if applicable) can be reviewed upon request
Reporting & Deliverables
• A complete calibration report including:
-
3-point temperature trend charts
-
Statistical analysis
-
Interpretation of deviations and improvement recommendations
• Suitable for: -
Internal QA documentation
-
Client audits (IQ/OQ/PQ)
-
ISO / regulatory compliance preparation
? READY Technical Advantages
• High-accuracy calibrated sensors with ISO/TAF traceability
• Experience with major shaking incubator brands (Thermo, Eppendorf, New Brunswick, Heidolph, etc.)
• Flexible calibration plan aligned with client SOPs and QA requirements
• Optional maintenance recommendations and troubleshooting support
For shaking incubator calibration (3-point temperature test with 1-hour recording), contact READY Biotech for on-site service and full documentation.