THERMO FRESCO 21 Centrifuge Motor Repair Record|READY Biotech
THERMO FRESCO 21 Centrifuge Motor Replacement|Service Record by READY
Content:
This repair case involved a THERMO Scientific FRESCO 21 benchtop centrifuge, reported by the client due to startup failure and unstable spinning. On-site inspection confirmed the root cause was a worn main motor.
The READY team replaced the motor with a compatible model, checked the belt and shaft alignment, and performed post-repair calibration, including RPM accuracy and timer precision tests, ensuring reliable operation.
Service items included:
-
Replacement of main motor
-
Inspection of belt and bearings
-
RPM accuracy testing (1,000 / 5,000 / 10,000 rpm)
-
Timer accuracy and spin stability test
-
Three trial runs to confirm no abnormal noise or vibration
Need repair service for your centrifuge? Contact READY for a professional on-site evaluation.