Vacuum Oven Temperature Comparison|9-Point Heat Test
Vacuum Oven Temperature Comparison|9-Point Heat Test
READY Biotech provides 9-point temperature distribution testing and panel-to-standard temperature comparison for laboratory vacuum ovens.
This service focuses on real-time evaluation after temperature stabilization—no data logging and no temperature curve reporting.
01|9-Point Heat Distribution Test
Nine sensors are positioned inside the chamber (Top/Middle/Bottom × Left/Center/Right) to evaluate:
-
Temperature uniformity
-
Hot and cold zones
-
Difference between set and actual temperature
02|Panel Temperature vs Standard Sensor Comparison
Once the oven reaches stable temperature, our engineer performs an immediate comparison:
-
Panel Display Temperature
vs -
Traceable Standard Temperature Sensor
This verifies:
-
Any temperature deviation (ΔT)
-
Whether adjustment or maintenance is required
-
Actual operating accuracy of the oven
03|Final Deliverables
You will receive:
-
9-point distribution results
-
Real-time panel vs standard comparison values
-
Recommendations if deviations are detected
☎︎ Tel: 02-8228-0131
✉️ Email: Ready_lab@outlook.com